Alignment and resolution studies of a MARS CT scanner A. Gongadze, A. S. Zhemchugov, P. Butler [et.al.]

Contributor(s): Gongadze, A | Butler, P | Bell, Steven J | Chelkov, G. A | Kozhevnikov, D. A | Potrap, I. N | Demichev, M. A | Smolyanskiy, P. I | Abramishvili, R | Kotov, S. A | Butler, A | Zhemchugov, A. SMaterial type: ArticleArticleSubject(s): пространственное разрешение | сканеры | томографияGenre/Form: статьи в журналах Online resources: Click here to access online In: Physics of particles and nuclei letters Vol. 12, № 5. P. 725-735Abstract: This paper describes a procedure of alignment of MARS CT scanner with cone beam geometry. The position of the xray source relative to the detector, the tilt of the detector are measured during alignment. The mechanical precision of gantry movement is determined. Significant improvement in the quality of image reconstruction has been achieved as a result of the calibration. Spatial resolution (PTF) of the scanner is confirmed to be about 30 μm.
Tags from this library: No tags from this library for this title. Log in to add tags.
No physical items for this record

Библиогр.: 5 назв.

This paper describes a procedure of alignment of MARS CT scanner with cone beam geometry. The position of the xray source relative to the detector, the tilt of the detector are measured during alignment. The mechanical precision of gantry movement is determined. Significant improvement in the quality of image reconstruction has been achieved as a result of the calibration. Spatial resolution (PTF) of the scanner is confirmed to be about 30 μm.